It can be used indoors as well as outdoors. For safety reasons, the instrument must only be operated within the permissible process conditions. You can find detailed information on the process conditions in chapter " Technical data " of the operating instructions or on the type label. Hence make sure before mounting that all parts of the instrument exposed to the process are suitable for the existing process conditions. Take note of chapter " Mounting instructions " for the recommended distance to the wall.
The emitted radar impulses of the radar sensor are electromagnetic waves. The polarisation is the direction of the electrical wave component.
By turning the instrument in the mounting strap, the polarisation can be used to reduce the effects of false echoes. The position of the polarisation is in the middle of the type label on the instrument.
In vessels with conical bottom it can be advantageous to mount the sensor in the centre of the vessel, as measurement is then possible down to the bottom.
Do not mount the instruments in or above the filling stream. Make sure that you detect the product surface, not the inflowing product. For socket mounting, the socket should be as short as possible and its end rounded.
This reduces false reflections from the socket. With threaded socket, the antenna end should protrude at least 5 mm 0. The socket end should be smooth and burr-free, if possible also rounded. When mounting on longer sockets, we recommend carrying out a false signal suppression see chapter " Parameter adjustment ".
You will find recommended values for socket heights in the following illustration or the table. The values come from typical applications. Deviating from the proposed dimensions, also longer sockets are possible, however the local conditions must be taken into account.
Vessel installations, such as e. Make sure when planning your measuring point that the radar sensor has a "clear view" to the measured product. In case of existing vessel installations, a false signal suppression should be carried out during setup.
If large vessel installations such as struts or supports cause false echoes, these can be attenuated through supplementary measures. Small, inclined sheet metal baffles above the installations scatter the radar signals and prevent direct interfering reflections. Detailed project planning data can be found at the channel manufacturers and in the technical literature. Welcome to ManualMachine. We have sent a verification link to to complete your registration.
Log In Sign Up. Forgot password? Enter your email address and check your inbox. Please check your email for further instructions.
Enter a new password. Vegapuls C Contents Contents 1 About this document Contents The respective scope of delivery results from the order specification. It is suitable for liquids and solids in virtually all industries, particularly in the water and waste water industry. Functional principle The instrument emits a continuous, frequency-modulated radar signal through its antenna. The emitted signal is reflected by the medium and received by the antenna as an echo with modified frequency.
The frequency change is proportional to the distance and is converted into the filling height. Its capacity to handle normal loads during transport is assured by a test based on ISO The packaging consists of environment-friendly, recyclable card- board. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Transport Transport must be carried out in due consideration of the notes on the transport packaging. Nonobservance of these instructions can cause damage to the device. Transport inspection The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or con- cealed defects must be appropriately dealt with.
Storage Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside. Ambient conditions Process conditions Mounting bracket Polarisation. Note: For safety reasons, the instrument must only be operated within the permissible process conditions.
If the sensor is installed in the center of dished or round vessel tops, multiple echoes can arise. However, these can be suppressed by an appropriate adjustment see chapter " Setup ". If you cannot maintain this distance, you should carry out a false signal suppression during setup. This applies particularly if buildup on the vessel wall is expected. In such cases, we recommend repeating the false signal suppression at a later date with existing buildup. Inflowing medium Mounting socket. Note: When mounting on longer sockets, we recommend carrying out a false signal suppression see chapter " Parameter adjustment ".
This ensures that the interfering reflections from the agitators are saved with the blades in different positions. If foams lead to measurement errors, you should use the biggest pos- sible radar antennas or sensors with guided radar. The following examples serve as an overview for flow measurement.
You can only view or download manuals with. Sign Up and get 5 for free. Upload your files to the site. You get 1 for each file you add. Get 1 for every time someone downloads your manual. Buy as many as you need. View and download manuals available only for. Register and get 5 for free. Upload manuals that we do not have and get 1 for each file. Get 1 for every download of your manual.
Buy as much as you need. Information sheet " Documents and software " with:. Information sheet " PINs and Codes " with:. Information sheet " Emergency unlock codes " with:. Optional instrument features are also described in this operating. The respective scope of delivery results from the. This operating instructions manual applies to the following instrument. The type label contains the most important data for identification and.
It is suitable for liquids and solids in virtually all industries, particularly. The instrument emits a continuous, frequency-modulated radar signal.
The emitted signal is reflected by the medium. Devices with integrated Bluetooth module can be adjusted wirelessly. Your instrument was protected by packaging during transport.
The packaging consists of environment-friendly, recyclable card-. Transport must be carried out in due consideration of the notes on the. Nonobservance of these instructions can cause. The delivery must be checked for completeness and possible transit. Ascertained transit damage or con-. Up to the time of installation, the packages must be left closed and. Unless otherwise indicated, the packages must be stored only under. The permissible storage and transport temperatures can be found in.
Screwed flanges are available in different versions according to the. Welded sockets are used to connect the sensors to the process. Enter your e-mail address and we will send you a link to reset your password. Reset password Cancel. We use cookies to continuously improve your experience at vega. Some are technically necessary to be able to provide you with our services.
Others help us to optimally adapt the content of our site to your requirements. By clicking on "Accept all", you agree to the use of all cookies. You can choose which cookies you want to accept by clicking on "Adjust settings".
You can find more information on this in our privacy policy. Select several products to compare with each other. Highlight multiple downloads to conveniently load these as Zip file. Application area The extended selection of process fittings and materials allows use in extreme temperature and pressure ranges. The sensor is ideal for level measurement in high silos, large bunkers, stone crushers and in the furnace.
Through the wide temperature and pressure range, the sensor can be used universally and enables a simple planning. Your benefit High plant availability, because wear and maintenance-free Maintenance-free operation through non-contact measuring principle Reliable measurement independent of vapour, dust and noise.
Product comparison Select several products to compare with each other. Technical specifications. Configure product. Level monitoring of bulk solids — a tough instrument for tough environments. Other applications Blast furnace. Level measurement in the blast furnace To the application. Sintering plant. Level measurement in the sintering plant To the application. Measurement of waste layer thickness and air flow in the incinerator To the application.
Pyrolysis chamber and ash container. Level measurement in the pyrolysis chamber and in the ash discharger To the application. Measuring range - Distance.
Process temperature. Process pressure. Beam angle. Materials, wetted parts. Alloy C22 2. Threaded connection. Flange connection. Seal material. Housing material. Stainless steel precision casting. Stainless steel electropolished. Protection rating. Profibus PA. Foundation Fieldbus. Ambient temperature. Explosion protection. Intrinsic safety "i". Flameproof enclosure "d". Dust explosion protection through housing "t". Ship approval. SIL qualification.
Software e. Drawing generator Generate 2D and 3D drawings for your specific configuration. Configure Now. Further downloads for this product All documents and software for this product can be found in our Download Center. Show all downloads.